Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips...
Philips Medical Systems (Cleveland) Inc began a recall of Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic... on . The reason FDA records is "Philips was informed that the detector buckets may experience weakness over time. As a result, this could cause the detector bucket to separate from the mounting plate of the gantry and eventually result in the detector...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0988-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Gamma camera systems: AXIS SPECT Systems & IRIX SPECT Systems, Model #: AXIS- 882130, IRIX- 882140. Philips Healthcare. Used to do nuclear medicine diagnostic imaging of various body organs.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Philips was informed that the detector buckets may experience weakness over time. As a result, this could cause the detector bucket to separate from the mounting plate of the gantry and eventually result in the detector falling.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 11/15/2011 Philips sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected product, the problem, the hazard involved, and the actions to be taken by users and the Philips. Customers are to place the letter in their User Documentation. Philips service will arrange a time to inspect customers' systems and conduct the appropriate field safety correction. Customers are to contact their local Philips representative if they need further support or information. In the US and Canada, customers can contact Philips Healthcare Customer Care Solutions Center at 1-800-722-9377.
- Quantity in commerce
- 795 units
- Distribution
- Worldwide Distribution -- USA, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PRO, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY and the territory of Puerto Rico; and countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Egypt, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, & United Kingdom.
- Product codes and lots
- System Code #, 882130, Serial #, 10, 97, 101, 103, 104, 107-112, 116, 117, 119-121, 123, 125, 127, 130, 132, 133, 135, 138, 142-144, 149-151, 153-158, 162-164, 166, 169-173, 180, 181, 185, 186, 189, 193-204, 214, 215, 218, 219, 221-223, 227, 229-231, 233-237, 240, 242, 244-246, 249-251, 260-262, 264, 267, 272, 273, 275, 277, 280, 281, 283, 285, 291-293, 295, 297-300, 305, 308-310, 312, 316, 319-322, 326-332, 334-336, 338-340, 346, 347, 349, 351, 352, 354, 355, 358-363, 368-370, 374, 377-379, 385, 387, 388, 390-396, 401-407, 411, 412, 416-419, 426-432, 434, 437, 438, 442, 443, 445-447, 449, 452, 457-462, 464, 465, 468, 471, 474, 480-485, 487, 490-493, 495, 496, 499, 502, 503, 505, 508, 509, 511, 513-518, 520, 521, 523, 529, 530, 534, 543-546, 548-560, 562, 563, 569-572, 574-576, 580, 582, 583, 585, 586, 590, 594, 595, 597, 602-607, 611, 612, 618-621, 624-628, 630, 635, 636, 640, 641, 647-650, 654, 656-658, 660-665, 668-673, 675, 679, 681-684, 686, 687, 690-695, 697-699, 701-708, 711-719, 723-729, 733-736, 738, 739, 743-746, 749-753, 755, 756, 759, 760, 762, 764-767, 771-773, 776-778, 780, 781, 786, 789, 790-795, 797, 799, 801, 803-806, 808, 809, 811-814, 818, 820-822, 825-827, 831-834, 836, 837, 839, 840, 841, 843-845, 848, 849, 851-855, 857, 858, 860-862, 865-869, 872-874, 876-881, 884-886, 889, 893, 894, 900-903, 907-922, 924, 927-930, 932, 934, 938-941, 943, 944, 946-948, 951, 953, 954, 958-963, 965-967, 970, 972-980, 982, 983, 985, 986, 991-994, 996-1000, 1003-1008, 1010-1013, 1016, 1017, 1019, 1022, 1023, 2007, 2008, 211037, 2008-0068, 555A, & 778A. System Code #882140, Serial # 102, 134, 141, 146, 147, 159, 160, 161, 167, 175, 176, 190, 209, 210, 216, 225, 226, 257, 259, 268-270, 274, 289, 290, 302, 304, 314, 315, 318, 323, 333, 341, 343, 344, 348, 350, 364, 366, 381, 386, 409, 414, 420, 421, 423, 436, 441, 455, 463, 469, 472, 473, 488, 489, 497, 498, 501, 531, 532, 536-539, 565, 566, 578, 588, 589, 591, 596, 598-600, 608, 610, 613, 615, 616, 622, 631, 632, 634, 637, 653, 666, 674, 676-678, 685, 689, 700, 709, 710, 731, 732, 741, 742, 748, 757, 763, 768, 774, 785, 802, 810, 816, 819, 824, 828, 829, 830, 838, 846, 856, 863, 864, 870, 871, 875, 884, 891, 892, 898, 904, 905, 931, 933, 936, 942, 945, 949, 950, 952, 955-957, 964, 968, 981, 989, 990, 995, 1001, 1002, 1009, 1015, 1018, & 1021. System Code #882130/40, Serial #184, 730, 122, 136, 254, 255, 284, 614, 696, 874, 906, & 984. System Code #882140/882130, Serial #113, 148, 152, 178, 179, 187, 188, 217, 238, 243, 247, 248, 278, 288, 345, 367, 371, 373, 380, 384, 389, 397, 397, 399, 408, 413, 415, 435, 439, 439, 440, 453, 477-479, 486, 504, 512, 525, 526, 528, 533, 535, 540, 547, 564, 567, 568, 577, 579, 581, 638, 639, 644, 645, 651, 652, 659, 667, 680, 747, 758, 815, 850, 859, 882, 883, 890, 896, 897, 899, 925, & 969.