Z-0876-2012

Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply...

CareFusion 303, Inc / initiated 2011-09-13 / Terminated / root cause: manufacturing process

CareFusion 303, Inc. began a recall of Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System on . The reason FDA records is "The recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0876-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
Product code
FRN Pump, Infusion
Reason for recall
The recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August 15, 2011 has a higher standby current resulting in the battery depleting prematurely.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Carefusion sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated September 13, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A Customer Reply Card was enclosed for customers to complete and return to the firm. Customers were instructed to call Carefusion Support Center at (888) 562-6018, 7 am-5 pm (Pacific) for any recall related questions.
Quantity in commerce
1039 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CT, FL, IA, IL, MI, MO, NJ, NY, OH, SC, SD, and TX. and the country of Australia.
Product codes and lots
The affected units have unique serial numbers for traceability.