Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply...
CareFusion 303, Inc. began a recall of Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System on . The reason FDA records is "The recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0876-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris PC unit model 8015. The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August 15, 2011 has a higher standby current resulting in the battery depleting prematurely.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Carefusion sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated September 13, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. A Customer Reply Card was enclosed for customers to complete and return to the firm. Customers were instructed to call Carefusion Support Center at (888) 562-6018, 7 am-5 pm (Pacific) for any recall related questions.
- Quantity in commerce
- 1039 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CT, FL, IA, IL, MI, MO, NJ, NY, OH, SC, SD, and TX. and the country of Australia.
- Product codes and lots
- The affected units have unique serial numbers for traceability.