Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45...
Abbott Vascular began a recall of Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0757-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Vascular
- Product
- Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55, 1011650-65, 1011651-45, 1011651-55. The RX Accunet EPS is indicated for use as a guide wire and embolic protection system.
- Product code
- NTE Temporary Carotid Catheter For Embolic Capture
- Reason for recall
- The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall thickness.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Abbott Vascular sent an "URGENT DEVICE RECALL" letter dated November 30, 2011 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to work with their local account representative to review their inventory, complete the attached Field Action Reconciliation/ Effectiveness Check Form and return any unused identified products to Abbott Vascular.
- Quantity in commerce
- 1253 units (962 units in the US & Puerto Rico)
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, China, Germany, Spain, France, United Kingdom, Italy, Pakistan, Poland, Russia, Slovakia, Taiwan, and South Africa.
- Product codes and lots
- Lot Number: 1081061, 1082561, 1091361, 1080561, 1081761, 1090261, 1091561, 1100461, 1080361, 1081661, 1090161, 1090861, 1092361, 1080861, 1082661, 1100761, 1081761, 1092861, 1082461, 1081161, 1082261, 1083061, 1081961, 1093061.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0757-2013 | Class II | Terminated | Voluntary: Firm initiated |