Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0934-2012

Imalux Niris OCT Imaging System Probe, Model Number 1300, Serial Number F464019, IMALUX Corporation...

Imalux Corporation / initiated 2011-12-22 / Terminated / root cause: manufacturing process

Imalux Corporation began a recall of Imalux Niris OCT Imaging System Probe, Model Number 1300, Serial Number F464019, IMALUX Corporation, Cleveland, OH The Niris Imaging System probe is used with... on . The reason FDA records is "During an FDA inspection, the firm was notified that the wrong tubing was used during the manufacture of the probe for the unit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0934-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Imalux Corporation
Product
Imalux Niris OCT Imaging System Probe, Model Number 1300, Serial Number F464019, IMALUX Corporation, Cleveland, OH The Niris Imaging System probe is used with the Niris Imaging System. The Niris Imaging System employs Optical Coherence Tomography (OCT), which can be used to construct high spatial resolution. The probe, is attached to the Imaging Console, and is used to direct light to and from the patient tissue. A small electromechanical scanning mechanism in the probe moves the optical beam laterally across the tissue surface while simultaneously acquiring in-depth backscattering profile at each lateral position. The probe is not labeled or sold as sterile.
Product code
NQQ System, Imaging, Optical Coherence Tomography (Oct)
Reason for recall
During an FDA inspection, the firm was notified that the wrong tubing was used during the manufacture of the probe for the unit.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Imalux Corp contacted their one customer via telephone on December 14, 2011. They informed the customer of the problem and requested the probe returned. The probe involved was returned and received by Imalux on December 22, 2011, and will be stored offsite in quarantine. Any further questions please call (216) 502-0755 ext. 301.
Quantity in commerce
One Unit
Distribution
Nationwide Distribution only to Texas.
Product codes and lots
Model Number 1300, Serial Number F464019