Philips MultiDiagnost Eleva FD Software release: PBL 6.1.1 Product Code 708037
Philips Healthcare Inc. began a recall of Philips MultiDiagnost Eleva FD Software release: PBL 6.1.1 Product Code 708037 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as... on . The reason FDA records is "The skin dose and skin dose rate values are calculated using an incorrect point of reference. Therefore, the AKR as displayed and stored in patient records is too low, although the dose as received by the patient is not...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0936-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips MultiDiagnost Eleva FD Software release: PBL 6.1.1 Product Code 708037 Product Usage: The Philips MultiDiagnost Eleva system is intended to be used as a multifunctional/universal system. General R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: Digestive system, Skeletal system, Urinary system, Reproductive system, Respiratory system, Circulatory system, and Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- The skin dose and skin dose rate values are calculated using an incorrect point of reference. Therefore, the AKR as displayed and stored in patient records is too low, although the dose as received by the patient is not affected.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare issued on January 9, 2012 a mandatory Field Safety Notice to all affected customers. The letter identified the affected product, problem description, hazard involved and actions that should be taken by the customer/user. The letter contains important information for the continued safe and proper use of the equipment. Customers were instructed to review the following information with all members of there staff and retain a copy with the equipment "Instruction for Use". For information or support concerning this notice contact your local Philips representative 866-767-2822.
- Quantity in commerce
- 28 units (US)
- Distribution
- Worldwide Distribution - Nationwide Distribution (USA) including the states of AR, AZ, CA, CO, FL, IA, KS, LA, MI, MO, NC, NY, OH, PA, TX, VA, VT, WA and WV
- Product codes and lots
- Software release: PBL 6.1.1