Z-1004-2012

CellTracks Analyzer II. A semi-automated fluorescence microscope used to enumerate fluorescently labeled...

Veridex, LLC / initiated 2011-08-17 / Terminated / root cause: software

Veridex, LLC began a recall of CellTracks Analyzer II. A semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immuno-magnetically selected and... on . The reason FDA records is "Control cell images may i intermix with Patient cell images while changing between the Control review screen and Patient review screen on the Cell tracks Analyzer II". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1004-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Veridex, LLC
Product
CellTracks Analyzer II. A semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immuno-magnetically selected and aligned. The system is for in-vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.
Product code
NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration
Reason for recall
Control cell images may i intermix with Patient cell images while changing between the Control review screen and Patient review screen on the Cell tracks Analyzer II.
Root cause tag
software
FDA root cause
Software design
Action
Veridex, LLC sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letter dated August 17, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter provides procedure sheets with step by step directions for customers to follow. Additionally, a Confirmation of Receipt form was enclosed for customers to complete and return to the firm. Contact Customer Technical Services at 1-877-837-4339 for questions regarding this notice.
Quantity in commerce
37 Units
Distribution
Worldwide Distribution.
Product codes and lots
K050145 (March 15, 2005); K060110 (March 16, 2006) Product code 9555