UV-Absorbing Posterior Chamber IOL Model: RS-55B, Serial Number: 1167121106, 1167131106, 1167141106...
Aaren Scientific, Incorporated began a recall of UV-Absorbing Posterior Chamber IOL Model: RS-55B, Serial Number: 1167121106, 1167131106, 1167141106, 1167151106, 1167161106, 1167171106, 1167181106... on . The reason FDA records is "The recall was initiated because Aaren Scientific has confirmed that a specific group of lenses were shipped without a complete expiry date showing on the label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1078-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aaren Scientific Inc.
- Product
- UV-Absorbing Posterior Chamber IOL Model: RS-55B, Serial Number: 1167121106, 1167131106, 1167141106, 1167151106, 1167161106, 1167171106, 1167181106, 1167191106, 1167201106, 1167211106, 1167221106, 1167231106, 1167241106, 1167251106, 1167261106, 1167271106, 1167281106, 1167291106, 1167301106, 1167311106, 1167321106, 1167331106, 1167341106, 1167351106, 1167361106, 1167371106, 1167381106, 1167391106, 1167401106, 1167411106, 1167421106, 1167431106, 1167441106, 1167451106, 1167461106, 1167471106, 1167481106, 1167491106, 1167501106, 1167511106, 1167521106, 1167531106, 1167541106, 1167551106, 1167561106, 1167571106, 1167581106, 1167591106, 1167601106, 1167611106, 1167621106, 1167631106, 1167641106, 1167651106, 1167661106, 1167671106, 1167681106, 1167691106, 1167701106, 1167711106, 1167721106, 1167731106, 1167741106, 1167751106, 1167761106, 1167771106, 1167781106, 1167791106, 1167801106, 1167811106, 1167821106, 1167831106, 1167841006, 1167851106, 1167861106, 1167871106, 1167881106, 1167891106, 1167901106, 1167911106, 1167921106, 1167931106, 1167941106, 1167951106, 1167961106, 1167971106, 1167981106, 1167991106, 1168001106, 1168011106, 1168021106, 1168031106, 1168041106, 1168051106, 1168061106, 1168071106, 1168081106, 1168091106, 1168101106, 1168111106, 1168121106, 1168131106, 1168141106, 1168151106, 1168161106, 1168171106, 1168181106, 1168191106, 1168201106, 1168211106, 1168221106, 1168231106, 1168241106, 1168251106, 1168261106, 1168271106, 1168281106, 1168291106, 1168301106, 1168311106, 1168321106, 1168331106, 1168341106, 1168351106, 1168361106, 1168371106, 1168381106, 1168391106, 1168401106, 1168411106, 1168421106, 1168431106, 1168441106, 1168451106, 1168461106. The device is indicated for primary implantation in the capsular bag of the eye for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed.
- Product code
- HQL Intraocular Lens
- Reason for recall
- The recall was initiated because Aaren Scientific has confirmed that a specific group of lenses were shipped without a complete expiry date showing on the label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Incorrect or no expiration date
- Action
- Aaren Scientific sent a recall communication letter dated August 2, 2011, August 3, 2011, and August 30, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Aaren Scientific is requesting that you return the lenses and we will rework the product and return them back to you. Please let me know and I will issue an RGA number and inform you of our FedEx account number to reference. In the meantime, we are conducting an internal investigation regarding this issue. Once this has been closed out then we will notify you of our findings. For further questions please call (909) 937-1033 ext. 108.
- Quantity in commerce
- 135 units
- Distribution
- Worldwide Distribution - USA including CA and TN, and the countries of India and Lebanon.
- Product codes and lots
- Lot Number: 110617