Z-1052-2012

Leica Bond Ready-to-Use Primary Antibody CD2 (11F11); 7 mL vial; an in vitro diagnostic primary antibody for...

Leica Microsystems, Inc. / initiated 2012-01-10 / Terminated / root cause: labeling and UDI

Leica Microsystems, Inc. began a recall of Leica Bond Ready-to-Use Primary Antibody CD2 (11F11); 7 mL vial; an in vitro diagnostic primary antibody for use as part of an antibody panel for the... on . The reason FDA records is "The primary antibodies to CD2 (clone 11F11) and CD7 (clone LP15) have been adulterated". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1052-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Leica Bond Ready-to-Use Primary Antibody CD2 (11F11); 7 mL vial; an in vitro diagnostic primary antibody for use as part of an antibody panel for the characterization of T cell disorders; Catalog No. PA0271; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom. For the qualitative identification by light microscopy of human CD2 molecule in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
The primary antibodies to CD2 (clone 11F11) and CD7 (clone LP15) have been adulterated.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Leica Microsystems, Inc. sent a Medical Device Field Notification letter dated January 4, 2012, via first class mail on January 10, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of the product and to return the product to the firm. Customers who have received the affected lot numbers of the listed products were requested to return any unused/part used reagents to the Buffalo Grove, IL, location for replacement. The accounts were asked to complete the attached Acknowledgement Form, acknowledging receipt and understanding of the letter, listing the lot numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the account's Leica Representative. For further questions please call (847) 405-0123
Quantity in commerce
114 - 7 mL vials
Distribution
Worldwide Distribution -- USA (nationwide) and the country of Canada.
Product codes and lots
Catalog No. PA0271, lot numbers 08173, 11855 and 14784