Z-0946-2012

Brilliance Big Bore Model #728243 and #728244, Philips Medical Systems. Computed Tomography X-Ray System...

Philips North America LLC / initiated 2011-12-13 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance Big Bore Model #728243 and #728244, Philips Medical Systems. Computed Tomography X-Ray System intended to produce of x-ray transmission data taken... on . The reason FDA records is "The Block and Multi Leaf Collimator (MLC) shape is not propagated to the beam that is modified by the Oppose operation. The contours can appear with the incorrect color in the Beam's Eye View Viewer, resulting in organs...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0946-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance Big Bore Model #728243 and #728244, Philips Medical Systems. Computed Tomography X-Ray System intended to produce of x-ray transmission data taken at different angles and planes.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The Block and Multi Leaf Collimator (MLC) shape is not propagated to the beam that is modified by the Oppose operation. The contours can appear with the incorrect color in the Beam's Eye View Viewer, resulting in organs being incorrectly placed or incorrectly shaped.
Root cause tag
software
FDA root cause
Software design
Action
On 12/13/2011, the firm sent Field Safety Notices, entitled URGENT - Medical Device Correction, to their customers. The letter identified the affected product, the problem, the hazard involved, and the actions to be taken by customers and the firm. Customers were to follow the instructions provided in order to address or avoid the issues. Philips will be installing a software update to address the identified issues free of charge. If customers need further information or support in regards to these issues, they are to contact their local Philips representative or local Philips Healthcare office. In North America and Canada, customers can contact the Customer Care Solutions Center at 1-800-722-9377.
Quantity in commerce
142 Units
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, CT, FL, IA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NH, NM, NY, OH, OR, PA, PR, TN, TX, VA, & WA and countries of AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, DENMARK, DOMINICAN REPUBLIC, FRANCE, GABON, GERMANY, HUNGARY, ISRAEL, ITALY, JORDAN, LUXEMBOURG, MALI, MEXICO, NETHERLANDS, OMAN, SLOVAKIA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, UKRAINE, UNITED ARAB EMIRATES, & UNITED KINGDOM.
Product codes and lots
Model # 728243, Serial # 7006, 7009, 7060, 7068, 7081, 7095, 7122, 7127, 7133, 7150, 7154, 7185, 7233, 7245, 7278, 7285, 7298, 7337, 7395, 7400, 7450, 7465, 7468, 7506, 7518, 7520, 7521, 7523, 7524, 7528, 7545, 7547, 7560, 7568, 7569, 7573, 7578, 7581, 7584, 7591, 7593, 7597-7599, 7602, 7611, 7616, 7617, 7618, 7620, 7622-7635, 7637-7650, 7652, 7655-7659, 7661-7663, 7665-7667, 7671-7675, 7677-7682, 7684-7687, 7689-7691, 7693, 7695, 7696, 7698-7707, & 75000.  Model # 728244, Serial # 7543, 7571, 7600, 7619, 7621, 7636, 7651, 7653, 7654, 7660, 7668-7670, 7676, 7683, 7688, 7694, 7697, 7708, & 7710.