AXIOM Artis and AXIOM Artis zee modular angiographic systems
Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis and AXIOM Artis zee modular angiographic systems Product Usage: Angiographic x-ray systems on . The reason FDA records is "There is a potential misalignment in the connection of the remote foot switch and the stationary unit. This potential malfunction may occur if either the wireless foot switch or the stationary unit is equipped with...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0927-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Artis and AXIOM Artis zee modular angiographic systems Product Usage: Angiographic x-ray systems
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- There is a potential misalignment in the connection of the remote foot switch and the stationary unit. This potential malfunction may occur if either the wireless foot switch or the stationary unit is equipped with improper firmware installed in the bluetooth interface performing the communication between foot switch and system. This is not a general issue but sporadic as a vendor unsystematical
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Siemens sent an Urgent Field Safety Notice letter on January 16, 2016, to all affected customers The letter identified the product, problem, potential risk, and actions to be taken. Customers were instructed to immediately pass this information to all those who need to be aware and instruct personnel accordingly. Siemens ask that if the device/equipment is no longer in the customers possession to forward this safety notice to the new owner of the device/equipment.
- Quantity in commerce
- 45
- Distribution
- Nationwide Distribution (USA) including the states of: CA, CT, FL, GA, HI, IL, IN, KY, ME, MD, MI, MS, MO, NH, NJ, NY, NC, OH, OR, TN, TX, UT, VA, WI and Puerto Rico.
- Product codes and lots
- Model numbers 5904433, 5904441, 5904649, 5904656, 5904466, 7007755, 7728350, 7412807, 7727717, 7728392, 7413078, 7555373, 7555365, 7555357, 7008605, 5917054, 10094135, 10094137, 10094139, 10094141, 10094142, 10094143, and 10280959.