Hamamatsu Corporation, Bridgewater NJ 08807 Microfocus X-ray Source L8601 series, L9121 series (90kV type...
Hamamatsu Corporation began a recall of Hamamatsu Corporation, Bridgewater NJ 08807 Microfocus X-ray Source L8601 series, L9121 series (90kV type MICROFOCUS X-RAY SOURCE), L7901 series, and L8031... on . The reason FDA records is "It was discovered that the Micro Focus X-ray Source could be experiencing software related problems that would allow the unit to exceed the maximum power setting that was set by the operator". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1224-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hamamatsu Corp.
- Product
- Hamamatsu Corporation, Bridgewater NJ 08807 Microfocus X-ray Source L8601 series, L9121 series (90kV type MICROFOCUS X-RAY SOURCE), L7901 series, and L8031 series (100kV type MICROFOCUS x-ray source
- Product code
- RBS Industrial X-Ray
- Reason for recall
- It was discovered that the Micro Focus X-ray Source could be experiencing software related problems that would allow the unit to exceed the maximum power setting that was set by the operator.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Hamamtsu Corporation sent a Warning letter dated December 22, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. CDRH approves your CAP under the following conditions: 1. Hamamatsu is able to work with your manufacturer customers to assure the revised manual and new label are provided to your customers' end users. 2. Hamamatsu informs CDRH if you receive information which indicates this control defect has occurred. Relevant information includes direct communication from customers or through observation of excessive wear on returned x-ray tubes. For questions regarding this recall call 908-231-0960, ext 2300.
- Quantity in commerce
- 100
- Distribution
- Worldwide Distribution-- USA (nationwide) and the country of Canada.
- Product codes and lots
- Series L8601, L9121, L7901, and L8031