Z-1044-2012

STA - System Control N + P

Diagnostica Stago, Inc. / initiated 2011-08-23 / Terminated / root cause: manufacturing process

Diagnostica Stago, Inc. began a recall of STA - System Control N + P Product Usage: The STA -System Control N+P kit provides a normal plasma and an abnormal plasma intended for use as two control... on . The reason FDA records is "Factor VIII quality control results are being recovered below the assigned ranges for STA System Control N+P". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1044-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diagnostica Stago, Inc.
Product
STA - System Control N + P Product Usage: The STA -System Control N+P kit provides a normal plasma and an abnormal plasma intended for use as two control levels for the following assays performed on analyzers of the STAbrand name suitable to these reagents: -the Reagent 1 (STA-System Control N)is for the following tests: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen (Clauss method), thrombin time (TT), Reptilase time, factors II, V, VII, VIII, IX, X, XI and XII, antithrombin (AT), protein C, protein S, plasminogen and antiplasmin. -the Reagent 2 (STA -System Control P) is for the tests as described above except thrombin time (TT) and Reptilase time. Do not use the STA -System Control N+P kit on STA Satellite for the control of the APTT performed with the STA-rTT kit.
Product code
GGN Plasma, Coagulation Control
Reason for recall
Factor VIII quality control results are being recovered below the assigned ranges for STA System Control N+P.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stago sent a recall letters via email dated August 23, 2011 to all affected customers . The letter identified the product, description of problem, and corrective actions to be taken. Customers were asked to immediately examine their inventory and follow the corrective actions provided. Customers were instructed to complete the response form attached to the letter. Because of incorrect manufacturer's instructions included in the initial letter another letter was emailed by the firm on August 30, 2011. For questions or technical assistance for updating the parameters please call the Diagnostica Stago 24 hour hotline at 1-800-725-0607.
Quantity in commerce
1259 units
Distribution
Worldwide Distribution and USA (Nationwide) and including Canada
Product codes and lots
K9483518 D029396 Lot #106775