Z-1212-2012
Philips Healthcare MD Eleva system. Multi-functional fluoroscopy, radiography and angiography systems that...
Philips North America LLC / initiated 2011-11-19 / Terminated / root cause: software
Philips Healthcare Inc. began a recall of Philips Healthcare MD Eleva system. Multi-functional fluoroscopy, radiography and angiography systems that can perform more specialized interventional... on . The reason FDA records is "It was discovered there was a software error where the air karma rate is not displayed correctly on the user interface". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1212-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare MD Eleva system. Multi-functional fluoroscopy, radiography and angiography systems that can perform more specialized interventional applications
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- It was discovered there was a software error where the air karma rate is not displayed correctly on the user interface.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A field correction action will be undertaken to correct the software error in affected systems worldwide. Your proposed action to bring the product into compliance with regulatory requirements is to institute a worldwide field correction to replace the software error in affected systems by updating the current software to version R6.1.2. The timeframe for completion of this field action will be 6 months post field action publication. Preventive actions: The software has been corrected moving forward in subsequent releases for these products. For further questions please call (978) 787-1501.
- Quantity in commerce
- 92 ( 24 Eleva 11s, and 68 Eleva FDs )
- Distribution
- Worldwide Distribution - USA (nationwide)
- Product codes and lots
- MultiDiagnost Eleva II - 708036, 708032, MultiDiagnost Eleva FD - 708037, 708034,