Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system
Siemens Medical Solutions USA, Inc began a recall of Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system on . The reason FDA records is "During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0940-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens MAGNETOM Aera/Skyra, model numbers 10432914 and 10432915. Nuclear magnetic resonance imaging system.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing necessary. The potential existed for a cable connection to become overheated, leading to the emission of smoke and possible material damage. Firm preventively checked the gradient cabl
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens Medical Solutions sent a "CUSTOMER SAFETY ADVISORY NOTICE" letter dated November 24, 2010 to all affected customers. The letter described the product, problem, and actions to be taken. Contact the firm at 610-219-4834 for questions regarding this recall.
- Quantity in commerce
- 128
- Distribution
- Nationwide Distribution-including the states of AK, AZ, CA, CT, DE, FL, GA, IL, IN, KS, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, and WI.
- Product codes and lots
- Serial numbers 45115, 41187, 41157, 41116, 45072, 45081, 41089, 41215, 41144, 41201, 45148, 41164, 41088, 41245, 41168, 41152, 45140, 45152, 45143, 45195, 41222, 41150, 45129, 45118, 45119, 41134, 41138, 45182, 45213, 41167, 41100, 52024, 45196, 45197, 41180, 41130, 45088, 41189, 41104, 41246, 41176, 45138, 15084, 41071, 45120, 45107, 41271, 45135, 45141, 41264, 45110, 41279, 41231, 41284, 41235, 45221, 45223, 41276, 41278, 41137, 41225, 41131, 41141, 41233, 45185, 41236, 41238, 45149, 41191, 45074, 45106, 45006, 45201, 41133, 41265, 41188, 45220, 41140, 41079, 45160, 45167, 41194, 45155, 45130, 41241, 41170, 45147, 41006, 45219, 45085, 45150, 45031, 41162, 45121, 45134, 45099, 45038, 45079, 41109, 41156, 45116, 41249, 45128, 45078, 45214, 41103, 45229, 41234, 45181, 52025, 52021, 45105, 45132, 41149, 45228, 45060, 45080, 45208, 41083, 45230, 45205, 41097, 41181, 45093, 45199, 45184, 45071, and 45111.