Z-1116-2012

Perkin Elmer Genetic Screening Processor (GSP)

Wallac Oy / initiated 2012-01-09 / Terminated / root cause: manufacturing process

PerkinElmer Life and Analytical Sciences, Wallac, OY began a recall of Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in... on . The reason FDA records is "GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1116-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wallac Oy
Product
Perkin Elmer Genetic Screening Processor (GSP) Product code: 2021-0010. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids.
Product code
KHO Fluorometer, For Clinical Use
Reason for recall
GSP instruments with defective washer modules may cause a false negative screening result being reported for T4, TSH and IRT. For 17-OHP, a false positive result may be generated if the failure occurs for a normal sample.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Perkin Elmer issued an "Urgent Medical Device Recall" letter dated 1/25/12 to all customers and informed them about the possible issue and requested permission for service personnel to check their instrument. The customers were instructed that the instrument could not be used and the washer module required replacement. Customers were told to expect to be contacted by the firm's local representative for further details.
Quantity in commerce
10 units
Distribution
Worldwide Distribution - USA, including the states of Florida, Ohio, Oregon, Texas and the countries of Brazil, Belgium, France, Germany, and the UK.
Product codes and lots
Serial numbers from 20210012 to 20210021