Z-1273-2012

IMMULITE 2000/IMMULITE 2000 XPi CK-MB L2KMB --- Catalog number: L2KMB2 (200 tests), L2KMB6 (600 tests)...

Siemens Healthcare Diagnostics Inc. / initiated 2012-01-25 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics began a recall of IMMULITE 2000/IMMULITE 2000 XPi CK-MB L2KMB --- Catalog number: L2KMB2 (200 tests), L2KMB6 (600 tests) --- Test code: CMB Color: Dark Blue. For in vitro... on . The reason FDA records is "Siemens Healthcare Diagnostics has observed that the median for the reference range with current kit lots of the CK-MB assay differs from the guidelines published in the Instructions for Use (IFU). This difference...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1273-2012
Status
Terminated
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMULITE 2000/IMMULITE 2000 XPi CK-MB L2KMB --- Catalog number: L2KMB2 (200 tests), L2KMB6 (600 tests) --- Test code: CMB Color: Dark Blue. For in vitro diagnostic use with the IMMULITE 2000 Analyzers --- for the quantitative measurement of creatine kinase isoenzyme MB (CK-MB) in heparinized plasma or serum, as an aid in patient management and the assessment of prognosis of myocardial infarction.
Product code
JHX Fluorometric Method, Cpk Or Isoenzymes
Reason for recall
Siemens Healthcare Diagnostics has observed that the median for the reference range with current kit lots of the CK-MB assay differs from the guidelines published in the Instructions for Use (IFU). This difference applies to all IMMULITE platforms.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an "URGENT DEVICE RECALL" letter dated January 2012 to all affected customers. The letter identifies the product , problem, and actions to be taken by the customers. Siemens recommends that customers reestablish their own medians and reference ranges, in line with the recommendations made in the IFU for these products. A Field Correction Effectiveness Check form was included for customers to complete and return to the firm. Customers with questions or who need additional information should contact their Siemens Technical Solutions Center.
Distribution
Worldwide Distribution.
Product codes and lots
kit lot 213, 214, 215