Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process
Elekta, Inc. began a recall of Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process on . The reason FDA records is "If a user, using an on-line correction protocol, clicks the accept correction button, an approval dialog box appears. If cancel is selected, the table correction values return to their initial values, but the image...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1066-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Synergy XVI 4.5 and R4.6 To be used as part of radiation therapy treatment process.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- If a user, using an on-line correction protocol, clicks the accept correction button, an approval dialog box appears. If cancel is selected, the table correction values return to their initial values, but the image panes continue to display the changed position.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- ELEKTA sent an Important Field Safety Notice (20001507042) dated January 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 770-670-2548
- Quantity in commerce
- 96 units
- Distribution
- Nationwide Distribution including AR, AZ, CA, CT, FL, GA, ID, IL, LA MD, MA, MI, MN, MO, NJ, NY, NC, ND, OH, OK, OR, PA, RI, TN, VA, WA, WI, and Puerto Rico
- Product codes and lots
- 152433, 152019, 151176, 152248, 152262, 152454, 105941, 151802, 152427, 151628, 151362, 152207, 151545, 151566, 152271, 152425, 152452, 152491, 151808, 152163, 152331, 151443, 151955, 152585, 152522, 151514, 151791, 151430, 151931. 105984, 151412, 152388, 151616, 151141, 152303, 151892, 151785, 152358, 152223, 152305, 152468, 151260, 152469, 152481, 152601, 152356, 152175, 152365, 152177, 152257, 151952, 152353, 152359, 152362, 152352, 152371, 152370, 151942, 151820, 151770, 151864, 151396, 152448, 151954, 151076, 151309, 152103, 152301, 151634, 151793, 152210, 152211, 151080, 152373, 152222, 152158, 152392, 151684, 152530, 152536, 151168, 151839, 151156, 151155, 151157, 152117, 151710, 152308, 152307, 152504, 151683, 151764, 151885, 152432, 151812, 152549.