REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet...
Biomet, Inc. began a recall of REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head... on . The reason FDA records is "Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1438-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head Metal Lock Pin is to be inserted into the anterior hole of the tibial bearing yoke to secure the OSS Axle in the correct position within the OSS Femoral component assembly.
- Product code
- KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
- Reason for recall
- Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin were evaluated and measured against print specifications.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Biomet sent an "URGENT MEDICAL DEVICE SAFETY ALERT" letter dated March 20, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to continue monitoring implanted patients for disassociation of the pin component and report to Biomet. The Alert included a complete listing of device invoice history and patient information (if supplied). Contact the firm for confirmation of receipt of the notice and for questions at 574-372-3983.
- Quantity in commerce
- 64 units distributed
- Distribution
- Distributed in California and Pennsylvania.
- Product codes and lots
- Catalog number: CP113456 with lot numbers: 593310, 743610, 750120, 815030, and 815040.