Z-1438-2012

REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet...

Zimmer GmbH / initiated 2012-03-20 / Terminated / root cause: design

Biomet, Inc. began a recall of REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head... on . The reason FDA records is "Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1438-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head Metal Lock Pin is to be inserted into the anterior hole of the tibial bearing yoke to secure the OSS Axle in the correct position within the OSS Femoral component assembly.
Product code
KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Reason for recall
Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin were evaluated and measured against print specifications.
Root cause tag
design
FDA root cause
Component design/selection
Action
Biomet sent an "URGENT MEDICAL DEVICE SAFETY ALERT" letter dated March 20, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to continue monitoring implanted patients for disassociation of the pin component and report to Biomet. The Alert included a complete listing of device invoice history and patient information (if supplied). Contact the firm for confirmation of receipt of the notice and for questions at 574-372-3983.
Quantity in commerce
64 units distributed
Distribution
Distributed in California and Pennsylvania.
Product codes and lots
Catalog number: CP113456 with lot numbers: 593310, 743610, 750120, 815030, and 815040.