DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of Depuy Mitek- Lupine Loop Plus Anchor w/ Panacryl Catalog Number: 210707 The Lupine Anchor System is indicated for use in soft tissue bone fixation on . The reason FDA records is "Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1164-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Mitek, Inc.
- Product
- Depuy Mitek- Lupine Loop Plus Anchor w/ Panacryl Catalog Number: 210707 The Lupine Anchor System is indicated for use in soft tissue bone fixation.
- Product code
- MAI Fastener, Fixation, Biodegradable, Soft Tissue
- Reason for recall
- Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- DePuy Mitek sent an Urgent Voluntary Product Recall letter dated January 27, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to immediately check all inventories to determine if they have any affected product. Customers were asked to refer to the instructions to report their inventory status, return and obtain replacement or credit for the product. For questions regarding this recall call 1-877-379-4871.
- Quantity in commerce
- 1757 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, AE, Argentina, Australia, Austria,Belgium,Brazil, Bahrain, Chile, China, Columbia,Cyprus, Check Republic, Germany, Denmark, Ecuador, Egypt, Spain, Finland, France, Finland, Great Britain, Guadeloupe, Hong Kong, Hungary,Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Luxembourg,Latvia, Martinique, Montenegro, Mexico, Maylasia, Netherlands,Norway, New Zealand, Oman, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Tunisia, Turkey,Uruguay, Venezuela, and South Africa
- Product codes and lots
- Lot Numbers: 3382594 3404938 3443265 3474050 3513370 3551904