Z-1193-2012

PointCare NOW CD4NOW Gold Reagent Pack, or Gold Pack, Part Number :30034 In Vitro Diagnostic : Assay reagent...

Pointcare Technologies, Inc. / initiated 2011-11-02 / Terminated / root cause: labeling and UDI

Pointcare Technologies, Inc. began a recall of PointCare NOW CD4NOW Gold Reagent Pack, or Gold Pack, Part Number :30034 In Vitro Diagnostic : Assay reagent intended for use on the PointCare NOW In Vitro... on . The reason FDA records is "Stability not assured at labeled temperature 2-42C". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1193-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Pointcare Technologies, Inc.
Product
PointCare NOW CD4NOW Gold Reagent Pack, or Gold Pack, Part Number :30034 In Vitro Diagnostic : Assay reagent intended for use on the PointCare NOW In Vitro Diagnostics Instrument to provide CD4 positive lymphocytes cell count
Product code
GKZ Counter, Differential Cell
Reason for recall
Stability not assured at labeled temperature 2-42C
Root cause tag
labeling and UDI
FDA root cause
Storage
Action
The firm, PointCare, issued a "Customer Service Advisory (CSA-00016)" dated November 2, 2011 to its customers to inform them of the new storage requirements of 2-30C as opposed to the previously claimed 2-42C. The letter describes the product, problem and action to be taken. The customers were instructed to correct the labeling on any Gold Reagent Packs and Gold Reagent and ensure that the Gold Reagent Packs and Gold Reagent they currently have in stock is stored within the 2-30C storage range. Future lots will be labeled with the updated storage temperatures. A member of our customer support team will provide additional information relative to this advisory. Should you have any questions or require further assistance, you may email us at [email protected] or call our offices directly at 1-508-281-6925.
Quantity in commerce
463 packs
Distribution
International only: including countries of : Abuja, Haiti, Indonesia, Natal, Nigeria, Senegal, South Africa, Uganda, and United Kingdom.
Product codes and lots
Lot Numbers:KG13230, KG13231, KG21520 and KG21521