Z-1144-2012

Tibiaxys Ankle Arthrodesis Plate Compression Guide, Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV...

Integra LifeSciences Corporation / initiated 2011-11-17 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Tibiaxys Ankle Arthrodesis Plate Compression Guide, Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH. Product Usage: The Compression guide... on . The reason FDA records is "During 6/30/2010 to 8/17/2011 the firm received five (5) incident complaints of breakage of the threaded part of the compression guide which potentially occurred when: 1) a compression is applied between talus and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1144-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Tibiaxys Ankle Arthrodesis Plate Compression Guide, Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH. Product Usage: The Compression guide is a re-usable instrument used in the implantation of the Tibiaxys product. The Tibiaxys system is intended for fixation of bone fractures or for bone reconstruction (arthrodesis, osteotomies, and fractures of ankle joint, distal tibia and fibula).
Product code
OYK Ankle Arthroplasty Implantation System
Reason for recall
During 6/30/2010 to 8/17/2011 the firm received five (5) incident complaints of breakage of the threaded part of the compression guide which potentially occurred when: 1) a compression is applied between talus and tibia. (Step 3 of the product description) or when 2) the surgeon screws the compression guide into the plate which most likely was attributed to improper surgical techniques used durin
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Integra sent a "Field Safety Notice" letter dated November 17, 2011 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review the Surgical Techinique Manual and follow the recommended guiidelines to avoid breakage. ensure that this information is distributed to their surgical teams, and fill out the enclosed "Field Safety Notice Acknowledgement and Return Form" and return to [email protected] or fax to 1-609-750-7999. For questions call 609-936-6932 or email at [email protected].
Quantity in commerce
168 units
Distribution
Worldwide Distribution - USA (nationwide) including the states of; AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA & WI. Product was also shipped to the following countries: Austria, Belgium, Canada, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Saudi Arabia, South Africa, Spain, Sweden, Switzerland & United Kingdom.
Product codes and lots
Model #159635, part # 159 635, Batch #: EAHZ, EAJ1, ELGV, ENQE & ENQH.