Orthofix rotating cutter used in spinal fusion procedures with the FORZA Spacer System. Catalog Number...
Orthofix, Inc. began a recall of Orthofix rotating cutter used in spinal fusion procedures with the FORZA Spacer System. Catalog Number 89-0452 (8mm Rotating Cutter), 89-0454 (10mm Rotating... on . The reason FDA records is "The 8mm and 10mm Rotating Cutters may fracture during use, specifically during disc space preparation. If an 8mm or 10mm Rotating Cutter does Fracture during use, it may result in a surgical delay estimated to be 30...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1037-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- Orthofix rotating cutter used in spinal fusion procedures with the FORZA Spacer System. Catalog Number 89-0452 (8mm Rotating Cutter), 89-0454 (10mm Rotating Cutter), and 89-0030 (FORZA Discectomy Case). Manual surgical instrument comprised of stainless steel and titanium.
- Product code
- max
- Reason for recall
- The 8mm and 10mm Rotating Cutters may fracture during use, specifically during disc space preparation. If an 8mm or 10mm Rotating Cutter does Fracture during use, it may result in a surgical delay estimated to be 30 minutes or greater due to preparing the disc space using other instruments or methods and/or retrieving the instrument fragment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Orthofix mailed recall letters to the distributors on 02/06/2012. Orthofix does not sell or consign these instruments to hospitals.
- Quantity in commerce
- 52
- Distribution
- Nationwide and Puerto Rico
- Product codes and lots
- Catalog Number 89-0452 (8mm Rotating Cutter) - Production Identification (Lot) Numbers 33056-ND25 and 33056-NE19; Catalog Number 89-0454 (10mm Rotating Cutter) - Production Identification (Lot) Numbers 33056-ND26 and 33056-NF02; Catalog Number 89-0030 (FORZA Discectomy Case); UPDATED INFORMATION AS OF 04/12/2012: The firm has revised to amend the affected lot information; lot number 34346-NH30 (PN 89-0545) has been added. At the time of recall initiation 3 units from lot 34346-NH30 were in distribution; all 3 units have since been returned to Orthofix. Recall notice stated all lots are to be returned so all other information remains unchanged.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
