Z-1375-2012

Access Immunoassay Systems Assay Protocol File (APF) Software versions for UniCel DxC 600i Systems. Part...

Siemens Healthcare Diagnostics Inc. / initiated 2008-09-11 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Access Immunoassay Systems Assay Protocol File (APF) Software versions for UniCel DxC 600i Systems. Part Number: A54378, UniCel DxC 600i APF, Part Number... on . The reason FDA records is "The recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1375-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay Systems Assay Protocol File (APF) Software versions for UniCel DxC 600i Systems. Part Number: A54378, UniCel DxC 600i APF, Part Number: A54379, UniCel DxC 600i AAF. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentration found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter sent a "PRODUCT CORRECTIVE ACTION" letter dated September 24, 2008 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers were instructed to complete and return the enclosed response form. Customers with questions regarding this notification were instructed to contact Technical Support at (800) 854-3633 or their local Beckman Coulter representative.
Quantity in commerce
2948 units total (2762 units in the US)
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Product codes and lots
Part Number: A54378, UniCel DxC 600i APF,  Lot Numbers: 722631, 823677. Part Number: A54379, UniCel DxC 600i AAF, Lot Numbers: 722630, 823675.