Synchron CX Immuno-Protein Cal, part number: 442840, other
Beckman Coulter Inc. began a recall of Synchron CX Immuno-Protein Cal, part number: 442840, other product code: CKA The Beckman Coulter SYNCHRON CX Immune Protein Calibrator Set, in conjunction with... on . The reason FDA records is "A recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1229-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Synchron CX Immuno-Protein Cal, part number: 442840, other product code: CKA The Beckman Coulter SYNCHRON CX Immune Protein Calibrator Set, in conjunction with SYNCHRON Immuno Protein Reagents, is intended for use on SYNCHRON CX Systems for the calibration of IgA, IgG, IgM, and Transferrin (TRF).
- Product code
- JIT Calibrator, Secondary
- Reason for recall
- A recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot number M901330.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent an Urgent Field Safety Notice letter dated August 24, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers with questions regarding the product correction action were instructed to contact their local Beckman Coulter Representative. For questions regarding this recall call 714-993-8916.
- Quantity in commerce
- 1,923 units total (1,882 units in US)
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada
- Product codes and lots
- Lot Number: M901330 and higher Other Lot numbers: M706000, M801451, M808220