Z-1229-2012

Synchron CX Immuno-Protein Cal, part number: 442840, other

Siemens Healthcare Diagnostics Inc. / initiated 2009-08-26 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Synchron CX Immuno-Protein Cal, part number: 442840, other product code: CKA The Beckman Coulter SYNCHRON CX Immune Protein Calibrator Set, in conjunction with... on . The reason FDA records is "A recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1229-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Synchron CX Immuno-Protein Cal, part number: 442840, other product code: CKA The Beckman Coulter SYNCHRON CX Immune Protein Calibrator Set, in conjunction with SYNCHRON Immuno Protein Reagents, is intended for use on SYNCHRON CX Systems for the calibration of IgA, IgG, IgM, and Transferrin (TRF).
Product code
JIT Calibrator, Secondary
Reason for recall
A recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot number M901330.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Beckman Coulter sent an Urgent Field Safety Notice letter dated August 24, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers with questions regarding the product correction action were instructed to contact their local Beckman Coulter Representative. For questions regarding this recall call 714-993-8916.
Quantity in commerce
1,923 units total (1,882 units in US)
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Product codes and lots
Lot Number: M901330 and higher  Other Lot numbers: M706000, M801451, M808220