Z-1225-2012

Smith & Nephew BIOSURE HA Interference Screw, 7 mm x 25 mm Catalog Number: 72201772. The BIOSURE HA...

Smith & Nephew, Inc. / initiated 2012-02-06 / Terminated / root cause: packaging

Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew BIOSURE HA Interference Screw, 7 mm x 25 mm Catalog Number: 72201772. The BIOSURE HA Interference Screw is indicated for fixation of... on . The reason FDA records is "Mislabeled: The device was labeled as a standard thread screw (P/N 72201772) on pouch and box, however the device contained in the packaging is a reverse thread screw (P/N 72201773)". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1225-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew BIOSURE HA Interference Screw, 7 mm x 25 mm Catalog Number: 72201772. The BIOSURE HA Interference Screw is indicated for fixation of bone-tendon-bone or soft tissue grafts during anterior/posterior cruciate ligament (ACL/PCL) reconstruction procedures.
Product code
HWC Screw, Fixation, Bone
Reason for recall
Mislabeled: The device was labeled as a standard thread screw (P/N 72201772) on pouch and box, however the device contained in the packaging is a reverse thread screw (P/N 72201773).
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Smith & Nephew Endoscopy sent an "URGENT-PRODUCT RECALL 1st NOTIFICATION" dated February 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The letter instructs customers to inspect their inventory and quarantine any affected product. For instructions on returning product, contact the Smith & Nephew Endoscopy Division Returns Group at 800-343-5717 ( 0ption 3) or email to [email protected]. Additionally, customers were asked to complete the bottom portion of the letter and to fax it to 1-508-261-3620. Contact the Smith & Nephew Endoscopy Division for questions regarding this notice at 1-508-261-3731.
Quantity in commerce
194 units
Distribution
Worldwide Distribution-USA (nationwide) and the states of AK, AZ, CA, CO, CT, IA , MD, MI, MN, MO, MS, NE, OH, OK, PA, SC, and VA and the countries of Australia, Canada, Germany, South Korea, Spain, Switzerland, Thailand, and United Kingdom.
Product codes and lots
Lot Number: 50397157