Siemens syngo.plaza with software version VA20B_HF04
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza with software version VA20B_HF04 Product Usage: Radiological image processing system on . The reason FDA records is "There is a potential malfunction when using syngo.plaza VA20B_HF04 with USB dongle and archive license. In syngo.plaza VA20B_HF04, there is a scenario where data in Short-Term Storage (STS) could reaming unprotected and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1127-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo.plaza with software version VA20B_HF04 Product Usage: Radiological image processing system
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- There is a potential malfunction when using syngo.plaza VA20B_HF04 with USB dongle and archive license. In syngo.plaza VA20B_HF04, there is a scenario where data in Short-Term Storage (STS) could reaming unprotected and not archived. this can happen when an archive server with a USB dongle-based license is enabled with auto archiving. If the disk fill level is reached which triggers auto deleti
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Customer Safety Advisory Notice letter dated January 20, 2012 to all affected consignees. The letter identified the affected product, problem, steps to take to avoid this issue, and actions to be taken. Customers were instructed to contact their local Siemens Service to plan for their software update and ensure that this safety advisory is placed in the system's instruction for use.
- Quantity in commerce
- 7
- Distribution
- Nationwide Distribution (USA) - CA, FL, GA, MO, NY and WA.
- Product codes and lots
- Model number 10592457 - serial numbers 100138, 100263, 100249, 100201, 100198, 100181 and 100146.