Z-1347-2012

Oridion Surestream CO2 Sampling Lines- Smart SureLine Plus O2 adult Part Number: 010981-03

Oridion Medical 1987 Ltd. / initiated 2011-12-01 / Terminated / root cause: design

Oridion Medical 1987 Ltd. began a recall of Oridion Surestream CO2 Sampling Lines- Smart SureLine Plus O2 adult Part Number: 010981-03 Product Usage: The intended use of SureStream sampling lines is to... on . The reason FDA records is "Surestream CO2 Sampling Lines-Connector may cause breakage of the microswitch inside the CO2 inlet". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1347-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Oridion Medical 1987 Ltd.
Product
Oridion Surestream CO2 Sampling Lines- Smart SureLine Plus O2 adult Part Number: 010981-03 Product Usage: The intended use of SureStream sampling lines is to convey a sample of a patients expired breath to a capnograph in order to measure CO2 partial pressure in intubated and non intubated patients.
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Surestream CO2 Sampling Lines-Connector may cause breakage of the microswitch inside the CO2 inlet
Root cause tag
design
FDA root cause
Device Design
Action
Oridion issued a Technical Note communication in March 2011 to field staff and end users globally. The Technical Note identified the affected product and provided instructions for user on the proper insertion of Surestream products. For questions contact your local distributor or Oridion at [email protected]
Quantity in commerce
70 boxes (1750 unts)
Distribution
USA Nationwide Distribution including the states of: AZ, CA, FL, HI, LA, MA, MO, NJ, NY, OH, OR, TX and WI.
Product codes and lots
Lot Numbers: M8209L10, M8262M10