Z-1195-2012

ACCESS Ostase Part Number: 37300 Used for the quantitative measurement of bone alkaline phosphatase (BAP)

Siemens Healthcare Diagnostics Inc. / initiated 2011-06-22 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of ACCESS Ostase Part Number: 37300 Used for the quantitative measurement of bone alkaline phosphatase (BAP) on . The reason FDA records is "The Access Ostase reagent lot 021281 shows instability resulting in falsely high patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1195-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ACCESS Ostase Part Number: 37300 Used for the quantitative measurement of bone alkaline phosphatase (BAP).
Product code
CIN Electrophoretic Separation, Alkaline Phosphatase Isoenzymes
Reason for recall
The Access Ostase reagent lot 021281 shows instability resulting in falsely high patient results.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A recall communication was initiated on 6/15/2011 with Beckman forwarding an "Urgent Product Correction" (PCA) letter to all their customers who purchased the Access Immunoassay Systems Ostase Reagent Kits. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to discontinue use of the Access Ostase reagent lot identified on the PCA letter and discard all remaining inventory. As stated in the Instructions for Use for Access Ostase, results should be interpreted in light of the total clinical presentation of the patient. All results generated with reagent lot number 021281 where patient results are higher than expected should be evaluated. The affected lot has been removed from Beckman's inventory, and alternate lots will be provided for all current and future replenishment orders or requests. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions can contact Customer Support Center at http://www.beckmancoulter.com/customersupport/support or call(800) 854-3633 in the US and Canada.
Quantity in commerce
318 units total (44 units in US)
Distribution
Worldwide Distribution -- US, including states of FL, IN, VA, CA, and KY and countries of China, Japan, and Panama.
Product codes and lots
Lot Number: 021281