Z-1099-2012

Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd...

Clarity Medical Systems / initiated 2012-02-02 / Terminated / root cause: software

Clarity Medical Systems Inc began a recall of Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic... on . The reason FDA records is "Still images extracted from video recordings may be mislabeled as to right and left eye". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1099-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Clarity Medical Systems
Product
Retcam3 & Retcam Shuttle with version 6.0 software. Clarity Medical Systems, Inc. 5775 W. Las Positas Blvd, Suite 200 Pleasanton, CA 94588 General ophthalmic imaging including retinal, corneal and external imaging. Photodocumentation of pediatric ocular diseases including retinopathy or prematurity (ROP). Screening of Type-2 re-threshold retinopathy of prematurity or treatment requiring ROP.
Product code
HKI Camera, Ophthalmic, Ac-Powered
Reason for recall
Still images extracted from video recordings may be mislabeled as to right and left eye.
Root cause tag
software
FDA root cause
Software design
Action
Clarity Medical Systems, Inc. sent a Safety Alert letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed the following : 1. Do not use the video mode until the software has been updated. 2. Install the software update using the enclosed CD. The instructions for installation are also enclosed. This software update will correct the anomaly. 3. If you have extracted still images from video recordings, review all potentially affected patient files to determine if any right eye images are mislabeled as left eye images and/or left eye images are mislabeled as right eye images. If any images are mislabeled, make the appropriate corrections in the annotation section of the affected file. Consider whether there is any diagnostic impact or need to re-irnage the patient. We apologize for the inconvenience. If you have questions or concerns, please contact Clarity at 800-215-6005.
Quantity in commerce
33 RetCam Shuttles and 20 RetCam 3 - total 53.
Distribution
Worldwide Distribution -- USA (nationwide) including the states of TN, MO, TX, FL, NJ and Puerto Rico and countries of Italy, Portugal, Czech Republic, Thailand, Chile, Oman, Hong Kong, Russia, Philippines, Sweden, Finland, Ecuador, Australia and Canada.
Product codes and lots
Model number RC3, part number 21-100500; Model number RCS, part number 20-000300.