Z-1194-2012

SYNCHRON Rheumatoid Factor (RF) Reagent, Part Number: 475136 A rheumatoid factor immunological test system is...

Siemens Healthcare Diagnostics Inc. / initiated 2011-06-20 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of SYNCHRON Rheumatoid Factor (RF) Reagent, Part Number: 475136 A rheumatoid factor immunological test system is a device that consists of the reagents used to... on . The reason FDA records is "The recall was initiated because Beckman has confirmed that the all in date lots of Synchron Rheumatoid Factor (RF) Reagent might generate false positive results or calibration failures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1194-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SYNCHRON Rheumatoid Factor (RF) Reagent, Part Number: 475136 A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues.
Product code
DHR System, Test, Rheumatoid Factor
Reason for recall
The recall was initiated because Beckman has confirmed that the all in date lots of Synchron Rheumatoid Factor (RF) Reagent might generate false positive results or calibration failures.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Beckman Coulter sent an Urgent Field Safety Notice/Product Correction letter dated June 20, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that Beckman Coulter is developing a new RF assay. In the meantime due to current issues related to shelf life and low end sensitivity, Beckman has decided to take the product off the market effective immediately. Beckman will keep the customers updated on their progress and advise when the new assay becomes available. Customers were instructed to complete and return the enclosed Response Form with in 10 days. Customers with questions regarding this notice can contact their local Beckman Sales representative or contact Customer Support Center at www.BeckmanCoulter.com/CustomerSupport/Support or call (800) 854-3633 in the US and Canada.
Quantity in commerce
22,081 units total (12,462 units in US)
Distribution
Worldwide Distribution - USA (nationwide) and worldwide to foreign countries: Australia, Belgium, China, Costa Rica, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macao, Mexico, Netherlands, New Zealand, Poland, Russian Federation, Serbia, Slovakia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Arab Emirates, and the United Kingdom.
Product codes and lots
All in-date lots