CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac...
Biosense Webster, Inc. began a recall of CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures on . The reason FDA records is "Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1230-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biosense Webster Inc
- Product
- CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Three Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customers are to contact their Biosense Webster representative. A Biosense Webster representative will contact customers to schedule an appointment for the replacement of the Backplane card. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. If customers have questions, they can contact their BWI representative.
- Quantity in commerce
- 134 units total (32 units in US)
- Distribution
- Worldwide Distribution -- US, Japan, China, Ireland, Germany, Italy, Slovenia, Indonesia, Poland, Spain, United Kingdom, Switzerland, Belgium, and Canada.
- Product codes and lots
- Catalog # FG-5400-00 Serial numbers: 11996, 11998, 11999, 12002, 12026, 12031, 12040, 12046, 12047, 12051, 12052, 12057, 12061, 12062, 12063, 12065, 12066, 12067, 12069, 12070, 12075, 12077, 12078, 12086, 12087, 12088, 12089, 12094, 12096, 12098, 12099, and 12111.