Z-1157-2012
SORIN GROUP, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004, USA, SDR S5 MAST ROLLER PUMP 85, Made...
LivaNova USA, Inc. / initiated 2012-02-16 / Terminated
Sorin Group USA, Inc. began a recall of SORIN GROUP, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004, USA, SDR S5 MAST ROLLER PUMP 85, Made in Germany, 1 UNITS PER CASE, CATALOG/PART No... on . The reason FDA records is "Touch screen may be unresponsive, inhibiting user input". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1157-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- SORIN GROUP, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004, USA, SDR S5 MAST ROLLER PUMP 85, Made in Germany, 1 UNITS PER CASE, CATALOG/PART No.: 50-80-70 The S5 System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six hours.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Touch screen may be unresponsive, inhibiting user input.
- FDA root cause
- Mixed-up of materials/components
- Action
- All affected US customers and distributors were notified of the Sorin S5 HLM Touch Screen Field Correction by certified mail on 02/15/12. The letter informed users of the potential for unresponsive S5 HLM touch screens and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to replace affected touch screens on site. Customers and distributors were also provided with a Response Form to confirm they had received, read and understood the Field Correction Notice. They were instructed to fax the completed form to (303) 467-6502. For questions regarding this recall call 303-467-6306.
- Quantity in commerce
- 0 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Belgium, Bulgaria, China, Denmark, Germany, Ecuador, Egypt, Finland, France, Georgia, Greece, United Kingdom, Hong Kong, Indonesia, Iraq, Iran, Israel, Italy, Japan, Jordan, Canada, British Columbia, South Korea, Kuwait, La Reunion, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Morocco, Mexico, Monaco, New Zealand, Netherlands, Norway, Oman, Austria, Phillippines, Poland, Portugal, Puerto Rico, Qatar, Serbia, Russia, Saudi Arabia, Sweden, Switzerland, Singapore, Spain, South Africa, Syria, Thailand, Czechoslavakia, Tunisia, Turkey, Hungary, Vietnam, and the United Arab Ermirates
- Product codes and lots
- Item number: 50-80-70 Serial numbers 50K00672 to 50K00688.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
