Access Ostase Reagent, Part Number: 37300, subsequent

Siemens Healthcare Diagnostics Inc. / initiated 2011-11-10 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Access Ostase Reagent, Part Number: 37300, subsequent product code: JIS. The Access Ostase Assay is used for the quantitative measurement of bone alkaline... on . The reason FDA records is "The recall was initiated because Beckman confirmed that the Access Ostase Calibrator and QC lots do not meet expiration date claims. The Access Ostase Reagent Kit lots identified were quality control tested and released...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1240-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Ostase Reagent, Part Number: 37300, subsequent product code: JIS. The Access Ostase Assay is used for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Pagets disease.
Product code
CIN Electrophoretic Separation, Alkaline Phosphatase Isoenzymes
Reason for recall
The recall was initiated because Beckman confirmed that the Access Ostase Calibrator and QC lots do not meet expiration date claims. The Access Ostase Reagent Kit lots identified were quality control tested and released using Access Ostase Calibrator and QC lots that contain the implicated raw material.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 9, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard all remaining inventory. In addition, a Response Form was enclosed for customers to complete and return. Contact Beckman's Customer Support Center at (800) 854-3633 for questions regarding this notice.
Quantity in commerce
8228 units total (1193 units in US)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, France, Georgia, Germany, Greece, Hong Kong, Hungary, India,Ireland, Italy, Japan, Jordan, Lebanon, Liechtenstein, Mexico, Morocco, Netherlands, New Zealand, Panama, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela.
Product codes and lots
Lot Numbers: Expiration dates: 011167 04/30/2011  017561 09/30/2011 014486 7/31/2011