Breeze Self-Adhesive Resin Cement, part number: N97F. The intended use is for cementation of porcelain...
Kerr Corporation began a recall of Breeze Self-Adhesive Resin Cement, part number: N97F. The intended use is for cementation of porcelain, resin, and metal-based inlays, onlays, crowns, bridges... on . The reason FDA records is "The recall was initiated because Pentron Clinical has confirmed that the Breeze Self-Adhesive Resin Cement exhibits faster than expected gel and set times. If the work time is exceeded when using this material, it could...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1100-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sybron Dental Specialties, Inc.
- Product
- Breeze Self-Adhesive Resin Cement, part number: N97F. The intended use is for cementation of porcelain, resin, and metal-based inlays, onlays, crowns, bridges, and posts.
- Product code
- EMA Cement, Dental
- Reason for recall
- The recall was initiated because Pentron Clinical has confirmed that the Breeze Self-Adhesive Resin Cement exhibits faster than expected gel and set times. If the work time is exceeded when using this material, it could possibly result in inadequate bond strength.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An "Urgent Medical Device Recall" letter was sent on 12/17/2010 via USPS 1st class mail to all their customers. The letter provides the customers with an explanation of the problem identified and the action to be taken. Customers were instructed to complete and fax back the enclosed recall return form. Customers with questions were instructed to call (800) 551-0283.
- Quantity in commerce
- 622 units total (12 units in US)
- Distribution
- Worldwide Distribution - Nationwide (USA) and the countries of Vietnam, Jordan,Turkey, and Italy.
- Product codes and lots
- Lot # 200466