Z-1268-2012

Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the...

Siemens Healthcare Diagnostics Inc. / initiated 2012-02-16 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in... on . The reason FDA records is "Automatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes diluted patient samples to under recover by approximately 50%". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1268-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems The ADVIA Centaur Vitamin D Total (Vit D) assay is for in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems. The ADVIA Centaur Vit Assay is intended as an aid in the determination of vitamin D insufficiency. The ADVIA Centaur Vitamin D Diluent 2-pack is used on board the ADVIA Centaur systems to perform automatic and system-performed dilutions fro Vitamin D assay.
Product code
MRG System, Test, Vitamin D
Reason for recall
Automatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes diluted patient samples to under recover by approximately 50%.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Siemens Healthcare Diagnostics, sent an "Urgent Device Recall Notice" on February 16, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to discontinue using the automatic and system-performed dilution features for Vitamin D; to discontinue the use of ADVIA Centaur VitD Diluent, 2-pack (REF 10494100); to manually dilute samples with values greater than the assay range of 150 ng/mL (375 nmoI/L), using the ADVIA Centaur VitD Diluent, 1 bottle (REF 10632114) as stated in the Vitamin D assay Instructions For Use (IFU); to review the contents of the recall notice with their Laboratory Director and retest the previous sample results generated using system- performed dilutions, and complete and return the confirmation fax-back form via fax to TECHNICAL SOLUTIONS CENTER at (302) 631-7597. If you have any questions or need additional information, please contact your local Technical Support Provider or Distributor or call (508) 668-5000.
Quantity in commerce
442 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Columbia, China, Denmark, Egypt, and Hong Kong.
Product codes and lots
Lot # 80230 Expiry Date 04 May 2012