Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The...
Siemens Medical Solutions USA, Inc began a recall of Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an... on . The reason FDA records is "This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1184-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an accessory used with the Linear Accelerator systems to deliver x-ray photon and electron radiation for the therapeutic treatment of cancer. This accessory shapes the radiation beam, per the leaf parameters, to minimize the amount of dose to outlying tissue.
- Product code
- IXI Block, Beam-Shaping, Radiation Therapy
- Reason for recall
- This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, SIEMENS, issued a "Customer Information" letter to its customers. The letter described the product, problem and actions to be taken. Siemens developed a new steel knob that complies with the latest standards; Siemens checked all ModuLeaf mMLC trolleys and if necessary refitted with the safety label and also arranged for an inspection and modification of the customers system- a Release of the safety update UI TH 026/11/S to distribute and hardware update commenced on February 7, 2007 by certified mail. If you have any questions, contact the Regulatory Specialist II at 925-602-8175.
- Quantity in commerce
- 46 active devices
- Distribution
- Worldwide distribution: USA (nationwide) including states of: CO, IA, KY, NJ, OH and TX; and countries of: Algeria, Austria, Belgium, Brazil, China, Egypt, France, Germany, Hungary, India, Italy, Japan, Kuwait, Malaysia, Mexico, P.R., Russian Fed., Saudi Arabia, South Africa, Spain, Turkey, and United Kingdom.
- Product codes and lots
- ModuLeaf, 110V/60Hz, part number 7334043; ModuLeaf, 220V/60Hz, part number 7334050.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.