Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1184-2012

Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The...

Siemens Medical Solutions USA, Inc. / initiated 2012-02-07 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an... on . The reason FDA records is "This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1184-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an accessory used with the Linear Accelerator systems to deliver x-ray photon and electron radiation for the therapeutic treatment of cancer. This accessory shapes the radiation beam, per the leaf parameters, to minimize the amount of dose to outlying tissue.
Product code
IXI Block, Beam-Shaping, Radiation Therapy
Reason for recall
This correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, SIEMENS, issued a "Customer Information" letter to its customers. The letter described the product, problem and actions to be taken. Siemens developed a new steel knob that complies with the latest standards; Siemens checked all ModuLeaf mMLC trolleys and if necessary refitted with the safety label and also arranged for an inspection and modification of the customers system- a Release of the safety update UI TH 026/11/S to distribute and hardware update commenced on February 7, 2007 by certified mail. If you have any questions, contact the Regulatory Specialist II at 925-602-8175.
Quantity in commerce
46 active devices
Distribution
Worldwide distribution: USA (nationwide) including states of: CO, IA, KY, NJ, OH and TX; and countries of: Algeria, Austria, Belgium, Brazil, China, Egypt, France, Germany, Hungary, India, Italy, Japan, Kuwait, Malaysia, Mexico, P.R., Russian Fed., Saudi Arabia, South Africa, Spain, Turkey, and United Kingdom.
Product codes and lots
ModuLeaf, 110V/60Hz, part number 7334043; ModuLeaf, 220V/60Hz, part number 7334050.