Z-1387-2012

Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human...

Roche Diagnostics GmbH / initiated 2012-03-12 / Terminated / root cause: manufacturing process

Roche Diagnostics Operations, Inc. began a recall of Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to... on . The reason FDA records is "An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1387-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return the attached fax form via fax to 1-888-912-8457 and file this letter for future references. If you have any questions about the information contained in this letter, please contact Roche Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-428-2336.
Quantity in commerce
739 kits
Distribution
Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, Algeria, Poland, Egypt, Qatar, Spain, Tunisia, France, Turkey, United Kingdom, Taiwan, India, Ukraine, Iran, Venezuela, Italy, South Africa, Jordan, and Armenia.
Product codes and lots
Elecsys Troponin I Immunoassay, 05094810160- lot 163176