Z-1159-2012

Siemens Ysio, AXIOM Luminos dRF, AXIOM Luminos Agile/TF, Uroskop Omnia Solid state x-ray imager (flat...

Siemens Medical Solutions USA, Inc. / initiated 2012-02-15 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Ysio, AXIOM Luminos dRF, AXIOM Luminos Agile/TF, Uroskop Omnia Solid state x-ray imager (flat panel/digital imager) on . The reason FDA records is "Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1159-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Ysio, AXIOM Luminos dRF, AXIOM Luminos Agile/TF, Uroskop Omnia Solid state x-ray imager (flat panel/digital imager)
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent Update Instruction XP057/11/S by mail on February 15, 2012, which releases software update package VC10D to remedy the described issues. For questions regarding this recall call 610-219-4834.
Quantity in commerce
58
Distribution
Nationwide Distribution including AK, CA, CO, FL, GA, IL, IN, IA, KY, MD, MA, MO, NE, NY, NC, OH, OR, PA, SC, TX, and UT.
Product codes and lots
Model numbers 10281013, 10094200, 10502200, 10094910