RayStation Version 2.0.0.15, Radiation Therapy Treatment Planning System. RaySearch Laboratories AB Sveavagen...
RAYSEARCH LABORATORIES AB began a recall of RayStation Version 2.0.0.15, Radiation Therapy Treatment Planning System. RaySearch Laboratories AB Sveavagen 25, SE-111 34 Stockholm Sweden... on . The reason FDA records is "An issue was found with the dose calculation in the RayStation 2.0, that could result in a potential overestimation of the dose. This could lead to the potential under-dosage of a patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1317-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation Version 2.0.0.15, Radiation Therapy Treatment Planning System. RaySearch Laboratories AB Sveavagen 25, SE-111 34 Stockholm Sweden. [email protected] Designed for treatment planning and analysis of radiation therapy.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- An issue was found with the dose calculation in the RayStation 2.0, that could result in a potential overestimation of the dose. This could lead to the potential under-dosage of a patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm decided to recall and sent Urgent Field Safety Notice Medical Device Correction letters to consignees on 12/06/2011. The letters described the issue, identified the affected product, actions to be taken by the user, a solution, and contact information. Users are to follow the two part workaround, and educate their staff and all users on the workarounds. The letter states that the issue will be resolved in the next version of RayStation. If further information is needed, users are to contact either Freddie Cardel, Director of Customer Support, RaySearch Americas, at 1-877-778-3849 or [email protected]; or Niclas Borglund, Director of Service, RaySearch Laboratories AB, at 46-8-5450-6130 or [email protected]
- Quantity in commerce
- 21 units (16 domestically and 5 internationally)
- Distribution
- Worldwide Distribution -- US, including the states of FL, MA, NJ, and NY and the country of the Netherlands.
- Product codes and lots
- Software version and build number is 2.0.0.15. Model number RayStation 2.0
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
