Brand Name: TempSpan CMT Temporary Cement, a temporary cement. Part Number: N69CA. The intended use of this...
Kerr Corporation began a recall of Brand Name: TempSpan CMT Temporary Cement, a temporary cement. Part Number: N69CA. The intended use of this device is to affix a temporary dental devices such... on . The reason FDA records is "The recall was initiated because Pentron Clinical has confirmed that the TempSpan Temporary Cement may have black plastic material on the surface of the paste, and it may set faster than specified". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1205-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sybron Dental Specialties, Inc.
- Product
- Brand Name: TempSpan CMT Temporary Cement, a temporary cement. Part Number: N69CA. The intended use of this device is to affix a temporary dental devices such as crowns or bridges to a tooth.
- Product code
- EMA Cement, Dental
- Reason for recall
- The recall was initiated because Pentron Clinical has confirmed that the TempSpan Temporary Cement may have black plastic material on the surface of the paste, and it may set faster than specified.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Pentron Clinical sent an Urgent Medical Device Recall letter dated April 20, 2011, to all affected customers. The letter provided the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to complete and fax back the enclosed recall form to 1-877-677-8844 in order to confirm receipt of the customer notification. Customers were instructed to contact Pentron Clinical Customer Service at (800) 551-0283, option 1, directly to handle the arrangements of a quick return and replacement. For questions regarding this recall call 714-516-7400.
- Quantity in commerce
- 627 units total (601 units in the US)
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, Philippines, Italy, Netherlands, and the UK.
- Product codes and lots
- Lot Number: 3748413