Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following...
Medtronic Vascular, Inc. began a recall of Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP... on . The reason FDA records is "Complaints received related to the inability/difficulty to release the proximal (FreeFlow) bare stent using the tip capture mechanism". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1284-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Talent Captivia Stent Graft System - FreeFlow configurations only. Model numbers affected have the following designation: TFXXXXXXXXCP and TBXXXXXXXXCP. Medtronic Inc. Santa Rosa, CA 95403. The Talent Thoracic Stent Graft System is designed to treat thoracic aortic aneurysms using an endovascular approach. When placed within the target lesion, the stent graft provides an alternative conduit for blood flow within a patients vasculature by excluding the lesion from blood flow and pressure.
- Product code
- MIH System, Endovascular Graft, Aortic Aneurysm Treatment
- Reason for recall
- Complaints received related to the inability/difficulty to release the proximal (FreeFlow) bare stent using the tip capture mechanism.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Medtronic, Inc. sent an Urgent Medical Device Safety Information letter dated February 2012, to all affected customers. The letter identified the product, the problem, and the corrective action taken by Medtronic. Medtronic updated the Instructions for Use (IFU) manual. For questions regarding this recall call 707-566-1110.
- Quantity in commerce
- 1523 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- TB2222C116CP, TF2424C116CP, TF2626C116CP, TF2828C116CP, TF3030C115CP, TF3232C115CP, TF3434C115CP, TF3636C114CP, TF3838C114CP, TF4040C114CP, TF4242C113CP, TF4444C113CP, TF4646C112CP, TB2222C161CP, TF2424C161CP, TF2626C161CP, TF2828C161CP, TF3030C160CP, TF3232C160CP, TF3434C160CP, TF3636C159CP; TF3838C159CP, TF4040C159CP, TF4242C158CP, TF4444C158CP, TF4646C157CP, TB2222C201CP, TF2424C201CP, TF2626C201CP, TF2828C201CP, TF3030C200CP, TF3232C200CP, TF3434C200CP, TF3636C199CP, TF3838C199CP, TF4040C199CP, TF4242C198CP, TF4444C198CP, TF4646C197CP.
FIRST ARTICLE / WEEKLY