Fresenius Optiflux F160NRe Hemodialyzer Catalog number: 0500316E Intended for patients with acute or chronic...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2012-03-06 / Terminated / root cause: manufacturing process

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Optiflux F160NRe Hemodialyzer Catalog number: 0500316E Intended for patients with acute or chronic renal failure when conservative therapy is judged... on . The reason FDA records is "Risk of internal blood leaks affecting the hemodialyzers". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1339-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Optiflux F160NRe Hemodialyzer Catalog number: 0500316E Intended for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Risk of internal blood leaks affecting the hemodialyzers.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Fresenius North America notified consignees on 3/6/12 by recall notification letter titled:Urgent FMCNA Optiflux F160NRe Hemodialyzer Recall, by certified mail with return receipt. The letter identified the affected product and described the reason for the recall. Customers were instructed to examine their stock immediately to determine whether they have any of the affected lot on hand. If customers have the affected lot, they have been instructed to immediately discontinue use and place all unused units in a secure area for return to Fresenius Medical Care North America. A Reply Response from is to be completed and returned. Customers have been instructed to contact their Fresenius Medical Care Customer Service Team for instructions on how to return the recalled product. Contact 1-800-323-5188.
Quantity in commerce
30,576 units
Distribution
Nationwide Distribution
Product codes and lots
Lot Number 11PU01012 Exp Date: 11/2014