Z-1178-2012

Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital...

Philips North America LLC / initiated 2011-04-01 / Terminated

Philips Healthcare Inc. began a recall of Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital radiography on . The reason FDA records is "It was discovered in some instances the detector did not reset as quickly as the user prompted a change form one image receptor (auxiliary) to the next, and then back again". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1178-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital radiography.
Product code
KPR System, X-Ray, Stationary
Reason for recall
It was discovered in some instances the detector did not reset as quickly as the user prompted a change form one image receptor (auxiliary) to the next, and then back again.
FDA root cause
Radiation Control for Health and Safety Act
Action
An Important Electronic Product Radiation Warning letter, dated February 28, 2012, was sent to customers. The letter identified the affected product, a description of the issue, the hazard involved, along with the actions to be taken and planned. Customers were to use one of the three work-a-rounds provided. The issue will be corrected with Field Change Order 70600064, which consists of a Software Update 4.0.2SP Service Pack installed on the system. If further information or support is needed, customers can contact Dominic Siewko at 978-659-7936.
Quantity in commerce
96
Distribution
Worldwide Distribution - Including USA.
Product codes and lots
706037