Z-1178-2012
Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital...
Philips North America LLC / initiated 2011-04-01 / Terminated
Philips Healthcare Inc. began a recall of Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital radiography on . The reason FDA records is "It was discovered in some instances the detector did not reset as quickly as the user prompted a change form one image receptor (auxiliary) to the next, and then back again". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1178-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare EasyDiagnost Eleva DRF 4.0 Stationary X-Ray System with DRF as an option for digital radiography.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- It was discovered in some instances the detector did not reset as quickly as the user prompted a change form one image receptor (auxiliary) to the next, and then back again.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- An Important Electronic Product Radiation Warning letter, dated February 28, 2012, was sent to customers. The letter identified the affected product, a description of the issue, the hazard involved, along with the actions to be taken and planned. Customers were to use one of the three work-a-rounds provided. The issue will be corrected with Field Change Order 70600064, which consists of a Software Update 4.0.2SP Service Pack installed on the system. If further information or support is needed, customers can contact Dominic Siewko at 978-659-7936.
- Quantity in commerce
- 96
- Distribution
- Worldwide Distribution - Including USA.
- Product codes and lots
- 706037