Z-1366-2013Class II

Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452

Depuy Orthopaedics Inc. / initiated 2012-03-07 / Terminated / root cause: packaging

DePuy Orthopaedics, Inc. began a recall of Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452 Product Usage: An acetabular liner is used in combination with other implants for... on . FDA classified it as Class II and gives the reason as "DePuy Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of one lot of the AltrX" liner because the product was incorrectly labeled on the box. The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1366-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452 Product Usage: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
DePuy Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of one lot of the AltrX" liner because the product was incorrectly labeled on the box. The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product enclosed was a 36IDX56OD Neutral Liner. Three ( 3 ) complaints were received in January 2012.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
DePuy Orthopaedics sent an URGENT INFORMATION- MEDICAL DEVICE RECALL NOTICE dated March 7, 2012 via email to all affected distributors and customers. The letter identified the affected product, problem and actions to be taken. Distributors and customers are instructed to cease further distribution or use or of recalled product and to return them to DePuy. The sales representatives are expected to aid customer in returning of recalled product. Distributors and sales representatives are to return the recalled product through the normal DePuy Returns process. Reconciliation form should be completed and returned to your DePuy Sales Representative or faxed to 574-372-7567. For questions about recall information contact the Manager of Customer Quality 574-372-7333.
Quantity in commerce
20 liners
Distribution
Worldwide Distribution - USA Nationwide including the state of VA and the country of Ireland.
Product codes and lots
product code1221-36-452, lot 187460.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1366-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1366-2013Class IITerminatedVoluntary: Firm initiated