Z-1474-2012

Philips eCareManager System, Model 865325, software versions 3.7, 3.7.1 and 3.8

Visicu, Inc. / initiated 2012-02-22 / Terminated / root cause: software

Visicu, Inc. began a recall of Philips eCareManager System, Model 865325, software versions 3.7, 3.7.1 and 3.8. Product Usage: The Philips eCareManager System software is intended for use in... on . The reason FDA records is "Three versions of software for the eCareManager failed to display anticoagulants on the medications screen. The problem arose due to an unexpected change in the National Drug Data File". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1474-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Visicu, Inc.
Product
Philips eCareManager System, Model 865325, software versions 3.7, 3.7.1 and 3.8. Product Usage: The Philips eCareManager System software is intended for use in data collection, storage, and clinical information management with independent bedside devices and ancillary systems that are connected either directly or through networks. The eCareManager system provides patient information and surveillance of monitored patients at the point of care location and at a remote, supplementary care location through wide area networking technology and dedicated phone lines and is solely intended for use in a hospital environment. It is not intended to be used in a home environment.
Product code
MSX System, Network And Communication, Physiological Monitors
Reason for recall
Three versions of software for the eCareManager failed to display anticoagulants on the medications screen. The problem arose due to an unexpected change in the National Drug Data File.
Root cause tag
software
FDA root cause
Software change control
Action
Philips notified all affected customers via a preliminary email message advising them of the issue. An URGENT - Field Safety Notice follow-up letter was mailed to all customers on February 22, 2012. This letter identified the affected product, described the problem and circumstances in which it can occur; recommend actions the customer should take to prevent risks for patients, and list the actions planned by Philips to correct the problem. For questions contact Philips eICU Customer Support 877-374-2872.
Quantity in commerce
12
Distribution
Nationwide Distribution (USA) including states of MA, MO, WI, AR, AZ, PA, SD, VA, and IN.
Product codes and lots
v3.7, v3.7.1, v3.8