Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment...
Aesculap Implant Systems began a recall of Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment of cervical spinal instability. The Quintex... on . The reason FDA records is "The firm has received complaints that the locking ring that sits in the head of the screw can come out of the screw head when engaged with the screwdriver if the screwdriver is improperly angled". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1455-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment of cervical spinal instability. The Quintex Cervical Screws are available in three different versions: semi-constrained, constrained, and dynamic. The screws all have a head diameter of 4.0mm and vary in length from 10mm to 22mm. Rescue screws are also available with a 4.5mm head diameter and lengths of 11-19mm.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- The firm has received complaints that the locking ring that sits in the head of the screw can come out of the screw head when engaged with the screwdriver if the screwdriver is improperly angled.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm initiated this recall by sending an "Important Recall Notification" letter to consignees on January 20, 2012. This letter instructed consignees not to use the product and requested that all stock be returned to the recalling firm promptly. Customers may contact 1-(866) 229-3002 for questions.
- Quantity in commerce
- 891
- Distribution
- Nationwide Distribution, including the states of CA, CO, FL, MO, and TX.
- Product codes and lots
- Semi-constrained -- SC400T - SC406T and SC490T - SC494T; Constrained -- SC500T - SC506T and SC590T - SC594T; Dynamic Screws -- SC600T - SC606T and SC690T - SC694T; all distributed lot numbers and serial numbers. The screws are part of the Quintex Cervical Plating System.