Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1455-2012

Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment...

Aesculap Inc / initiated 2012-01-24 / Terminated / root cause: manufacturing process

Aesculap Implant Systems began a recall of Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment of cervical spinal instability. The Quintex... on . The reason FDA records is "The firm has received complaints that the locking ring that sits in the head of the screw can come out of the screw head when engaged with the screwdriver if the screwdriver is improperly angled". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1455-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Quintex Cervical Plating System - Screws. The Quintex Cervical Plating System is intended for the treatment of cervical spinal instability. The Quintex Cervical Screws are available in three different versions: semi-constrained, constrained, and dynamic. The screws all have a head diameter of 4.0mm and vary in length from 10mm to 22mm. Rescue screws are also available with a 4.5mm head diameter and lengths of 11-19mm.
Product code
KWQ Appliance, Fixation, Spinal Intervertebral Body
Reason for recall
The firm has received complaints that the locking ring that sits in the head of the screw can come out of the screw head when engaged with the screwdriver if the screwdriver is improperly angled.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm initiated this recall by sending an "Important Recall Notification" letter to consignees on January 20, 2012. This letter instructed consignees not to use the product and requested that all stock be returned to the recalling firm promptly. Customers may contact 1-(866) 229-3002 for questions.
Quantity in commerce
891
Distribution
Nationwide Distribution, including the states of CA, CO, FL, MO, and TX.
Product codes and lots
Semi-constrained -- SC400T - SC406T and SC490T - SC494T; Constrained -- SC500T - SC506T and SC590T - SC594T; Dynamic Screws -- SC600T - SC606T and SC690T - SC694T; all distributed lot numbers and serial numbers. The screws are part of the Quintex Cervical Plating System.