Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1391-2012

EasyLink(TM) Informatics Systems Software Version 5.0 and Software Version 5.0 Service Packs 1 through 4 used...

Siemens Healthcare Diagnostics Inc. / initiated 2011-12-02 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of EasyLink(TM) Informatics Systems Software Version 5.0 and Software Version 5.0 Service Packs 1 through 4 used with the Dimension Vista(R) System. The... on . The reason FDA records is "QC results may be unexpectedly associated with a Pending Control. When this situation occurs, Pending QC Controls maybe be created or reused". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1391-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
EasyLink(TM) Informatics Systems Software Version 5.0 and Software Version 5.0 Service Packs 1 through 4 used with the Dimension Vista(R) System. The EasyLink(TM) Informatics System is classified as a data calculator and processing module for use with In-vitro diagnostics devices. The EasyLink(TM) is included with the Dimension Vista(R) System, and may be used with other Siemens analyzers, as a communications and connectivity workstation for integration with laboratory information system (LIS) networks. Product Quantity Distributed (Int) 787
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
QC results may be unexpectedly associated with a Pending Control. When this situation occurs, Pending QC Controls maybe be created or reused.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens Healthcare Diagnostic Inc, sent a Urgent Field Safety Notice dated December 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were asked to follow these instructions. The first option is Extending the "Minimum Inactive Days" to 999 days. The second option is to proactively monitor the system for unexpected Pending QC Controls. Long Term Action Plan: A corrective and preventive action (CAPA) investigation has been issued. Currently firm is actively investigating. Further questions please call (302) 631-0516.
Quantity in commerce
787
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of CANADA, FRANCE, GERMANY, SPAIN, PORTUGAL, ITALY, JAPAN, NORWAY, AUSTRALIA, AUSTRIA, NEW ZEALAND, BELGIUM, SWITZERLAND, NETHERLANDS, MALAYASIA, SLOVENIA, SOUTH KOREA, DENMARK, SAUDI ARABIA, SLOVAKIA, PUERTO RICO, INDONESIA and TAIWAN.
Product codes and lots
Catalog number 1000034805