Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models...
Medtronic Inc. Cardiac Rhythm Disease Management began a recall of Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153DRG and D154DRG. The implantable cardioverter... on . The reason FDA records is "A small percentage of EnTrust ICDs may not meet expected longevity or provide at least three months of device operation between the Elective Replacement Indicator (ERI) and End of Life (EOL) due to a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1250-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153DRG and D154DRG. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 55432 USA.
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- A small percentage of EnTrust ICDs may not meet expected longevity or provide at least three months of device operation between the Elective Replacement Indicator (ERI) and End of Life (EOL) due to a more-rapid-than-expected drop in battery voltage.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An "Urgent Medical Device Correction" letter was sent to physicians beginning 3/6/2012. The letter described the issue and provided patient management recommendations. Additionally the letter stated that ongoing updates to the product performance report will be posted on a Medtronic website.
- Quantity in commerce
- 68,804 devices worldwide
- Distribution
- Worldwide Distribution -- USA, and the countries of: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jamaica, Jordan, Kuwait, Lebanon, Macedonia, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Reunion, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Product codes and lots
- all serial numbers