Z-1283-2012
Philips Easy Diagnost Eleva DRF, Software release 4.0 Fluoroscopic image-intensified Stationary X-Ray System
Philips Medical Systems DMC GmbH / initiated 2012-03-09 / Terminated / root cause: software
Philips Medical Systems Gmbh, DMC began a recall of Philips Easy Diagnost Eleva DRF, Software release 4.0 Fluoroscopic image-intensified Stationary X-Ray System on . The reason FDA records is "Underexposed image due to lag in switching between auxiliaries". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1283-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips Medical Systems DMC GmbH
- Product
- Philips Easy Diagnost Eleva DRF, Software release 4.0 Fluoroscopic image-intensified Stationary X-Ray System.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Underexposed image due to lag in switching between auxiliaries
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated February 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The firm provided a work around for customers to use until a software update can be installed. Customers should contact their local Philips representative for questions regarding this notice.
- Quantity in commerce
- 88 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Germany, Indonesia, Malaysia, Saudi-Arabia, Singapore, and Thailand.
- Product codes and lots
- Software release 4.0