BioTron Sickle Cell Reagent Set, Screening test for the Presence of Hemoglobin S 2 x 100 ml Sickle Cell...

MichClone Associates, Inc. / initiated 2012-02-23 / Terminated / root cause: manufacturing process

MichClone Associates, Inc. began a recall of BioTron Sickle Cell Reagent Set, Screening test for the Presence of Hemoglobin S 2 x 100 ml Sickle Cell Buffer, 2 vials sickle Cell Reagent, Size 100 tests... on . The reason FDA records is "Two complaints received concerning the powder vial included in the 100 test kit observed producing pressure while sealed. When the powders are opened, the pressure from within the vial releases a small amount of powder...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1502-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MichClone Associates, Inc.
Product
BioTron Sickle Cell Reagent Set, Screening test for the Presence of Hemoglobin S 2 x 100 ml Sickle Cell Buffer, 2 vials sickle Cell Reagent, Size 100 tests, Store 2-8 C, BioTron Diagnostics Inc, Hemet California, USA . For the qualitative determination of positive or negative Sickle Cell Anemia present in Hemoglobin , for in vitro diagnostic use only.
Product code
GHM Test, Sickle Cell
Reason for recall
Two complaints received concerning the powder vial included in the 100 test kit observed producing pressure while sealed. When the powders are opened, the pressure from within the vial releases a small amount of powder into the air. This powder. if released near the face and inhaled can cause minor irritation resulting in sneezing and nasal congestion. Investigation found no affect on the perf
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Michclone Associates, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 15, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to identify affected lots, notify sub accounts, and complete and return the response form to arrange return of unused product and replacement . Forms can be returned by FAX 248(307)-9498, mail: Michclone Associates Inc. 680 Ajax Drive, Madison Heights, MI 48118 or Email: [email protected]. Questions and concerns should be addressed to Technical Support Department at 248-307-9494.
Quantity in commerce
250
Distribution
Nationwide Distribution.
Product codes and lots
Lot# 1187, expiration date: 7/13  Catalog No.25100, Powder Lot # 1171 (part of kit)